CRY334 Measurement microphone
Pressure field type; pre-polarization; frequency range: 3.15 Hz - 10 kHz; sensitivity: 50 mV/Pa; dynamic range: 15-146 dB
CRY334 Microphone is a 1/2'' pre-polarization pressure-field type measurement microphone that measures the actual sound pressure on the surface of the microphone diaphragm (including changes in the sound field due to the presence of the microphone itself).
Standard-compliant
Compliance with IEC61094-4: 1995 Measurement microphones - Part 4: Specifications for working standard microphones
Compliance with GBT20441.4-2006 Measurement microphones - Part 4: Specifications for working standard microphones
Technical specification of CRY334 |
---|
Sound field type | Pressure field type |
Sensitivity (@ 250 Hz ± 2 dB) | 50 mV/Pa |
Sensitivity (@ 250 Hz ± 2 dB) | -26 dB re 1V/Pa |
Frequency response (±2 dB) | 3.15Hz~10kHz |
Polarization voltage | 0V(pre-polarization) |
Typical capacitance | 17pF |
Upper limit of dynamic range (re. 20 uPa) | ≥146dB(distortion < 3%) |
Lower limit of dynamic range (re. 20 uPa) | ≤15dB(A) |
Range of operating temperature | -30℃~80℃ |
Temperature coefficient (-10 ℃ to 50 °C) | 0.012dB/℃ |
Static pressure coefficient | -0.01dB/kPa |
Relative humidity range (without condensation) | 0~90%RH |
Relative humidity coefficient | <0.1dB(0 - 90% RH without condensation) |
Pressure equalization method | Post-pressure equalization |
IEC 61094-4 Designation | WS2F |
Height (without protective grid) | 16.2mm |
Diameter (including protective grid) | 13.2mm |
Height (without protective grid) | 15mm |
Diameter (without protective grid) | 12.7mm |
Preamplifier with thread | 11.7mm-60UNS |
-
20 years industry experience
-
24 hours Technical Support
-
Professional engineers on line response
-
2 hours response
Solution within 2 work days -
All life Costs Maintenance