CRY319 Ear simulator/coupler
2 cc coupled cavity, resonant frequency = 22.4 kHz
CRY319 is a kind of coupled cavity with the volume of 2 cc in accordance with IEC 60318-5. It can be adapted to 1/2'' pressure-field measurement microphones according to IEC 61094-4 (WS2P), such as CRY372, CRY332, CRY334 and other measurement microphones. Hearing aids and headphones for air conduction are connected to the coupled cavity by ear inserts (e.g. ear molds or similar devices). The coupler has a resonant frequency of 22.4 kHz to avoid interference with measurements within the audible frequency range. This coupled cavity kit has a built-in CRY372 measurement microphone with the sensitivity of 12.5 mV/Pa and the noise floor of 20 dB(A).
-
High stability
-
High precision
-
Interference immunity
Standard-compliant
Compliance with IEC 60318-5:2006 Part 5: 2 cm³ coupler for the measurement of hearing aids and earphones coupled to the ear by means of ear inserts
Compliance with ANSI S3.7:2008 American National Standard Method for Coupler Calibration of Earphones
Compliance with IEC 61094-4:1995 Part 4: Specifications for working standard microphones
Compliance with GB/T 20441.4-2006 Part 4: Specifications for working standard microphones
Case Dimension Drawing of CRY319
CRY372 Measurement Microphone is a pre-polarization 1/2'' pressure-field type condenser microphone. It is mainly used for testing high-frequency response and high SPL acoustic signals. The upper limit of dynamic range is up to 160 dB. This microphone uses a titanium film as a diaphragm to ensure high stability and high accuracy in harsh environments.
CRY510 Preamplifier can be combined with a 1/2" measurement microphone for impedance conversion. With high input impedance, low output impedance, low noise, wide frequency range, and other characteristics, this preamplifier can be used in electroacoustic test systems or multi-noise detection analysis.
Case Dimension Drawing of CRY319
CRY372 Measurement Microphone is a pre-polarization 1/2'' pressure-field type condenser microphone. It is mainly used for testing high-frequency response and high SPL acoustic signals. The upper limit of dynamic range is up to 160 dB. This microphone uses a titanium film as a diaphragm to ensure high stability and high accuracy in harsh environments.
CRY510 Preamplifier can be combined with a 1/2" measurement microphone for impedance conversion. With high input impedance, low output impedance, low noise, wide frequency range, and other characteristics, this preamplifier can be used in electroacoustic test systems or multi-noise detection analysis.
CRY319 Ear Simulator / Coupled Cavity |
---|
Volume of coupled cavity | 2cc ± 2% |
Resonant frequency | 22.4kHz ± 2.5kHz |
Frequency Response (FR) | 125Hz~8kHz |
Sensitivity ±2 dB | 12.5mV/Pa |
Sensitivity ±2 dB | -38dB re 1V/Pa |
Polarization voltage | 0V(pre-polarization) |
Upper limit of dynamic range (re 20 uPa) | 160dB |
Lower limit of dynamic range (re 20uPa) | 20dB(A) |
Operating temperature | -30℃~80℃ |
Temperature coefficient (-10 ℃ - 50 ℃) | 0.005dB/℃ |
Relative humidity range | 0~90% (without condensation) |
Power Supply of Preamplifier | ICP(2~10mA) |
Connection Mode of Preamplifier | SMB coaxial plug/BNC(CRY510) |
Configuration checklist |
---|
CRY319 2 cc Coupled Cavity |
CRY372 1/2'' Pressure-Field Measurement Microphone |
CRY510 1/2'' Preamplifier |
BL5706 SMB - BNC cable / 1.5 m |
-
20 years industry experience
-
24 hours Technical Support
-
Professional engineers on line response
-
2 hours response
Solution within 2 work days -
All life Costs Maintenance